Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med DASH IV: (1 year) CAL VER /5500,34401 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 96 CONFIGURATION: HP 34401A ============================================================================= # # Source: # Astro-Med Calibration Verification Checklist # # Input LOpatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # Sub Astro-Med DASH Series Gain (1 year) CAL VER /5500 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- N P F V 1.003 ASK+ X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Turn unit ON. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature: 23 +/-2 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {GAIN SETTINGS} 1.008 DISP Press the SYSTEM key to access the DEFAULT submenu. 1.008 DISP Press the ZERO key in the DEFAULT submenu to 1.008 DISP center each trace to zero. 1.008 DISP Press the CALIBRATE key on the front panel. 1.009 DISP Set the range for each channel to .5 VFS 1.010 MEM2 =Channel 1 1.011 CALL Sub Astro-Med DASH Series Gain (1 year) CAL VER /5500 1.012 JMP 2.001 1.013 EVAL dummy 2.001 MEM2 =Channel 2 2.002 CALL Sub Astro-Med DASH Series Gain (1 year) CAL VER /5500 2.003 JMP 3.001 2.004 EVAL dummy 3.001 MEM2 =Channel 3 3.002 CALL Sub Astro-Med DASH Series Gain (1 year) CAL VER /5500 3.003 JMP 4.001 3.004 EVAL dummy 4.001 MEM2 =Channel 4 4.002 CALL Sub Astro-Med DASH Series Gain (1 year) CAL VER /5500 4.003 JMP 5.001 4.004 EVAL dummy 5.001 MESS 5.002 DISP Disconnect the 5500A and remove the recorder's top. 5.003 HEAD {PRECISION INTERNAL VOLTAGE REFERENCES} 5.004 DISP Connect the 34401A to U23 on the UUT A/D board as 5.004 DISP follows: 5.004 DISP [32] 34401A Input HI to pin 7 (precision reference) 5.004 DISP [32] 34401A Input LO to pin 5 (signal ground). 5.005 34401 2.5000V 0.0025U 2W 6.001 HEAD {POWER SUPPLY VOLTAGES} 6.002 DISP Connect the 34401A to connector P13/P2 on the UUT 6.002 DISP 31637-001 Power Supply board as follows: 6.002 DISP [32] 34401A Input HI to Red (+) 6.002 DISP [32] 34401A Input LO to Black (-) 6.003 34401 5.0V 0.5U 2W 7.001 DISP Connect the 34401A to P4 on the UUT follows: 7.001 DISP [32] 34401A Input HI to Grey (+) 7.001 DISP [32] 34401A Input LO to Black (-) 7.002 34401 24.0V 2.4U 2W 8.001 DISP Connect the 34401A to connector P2 on the UUT 8.001 DISP 31635-001 Power Supply board as follows: 8.001 DISP [32] 34401A Input HI to Orange (+) 8.001 DISP [32] 34401A Input LO to Black (-) 8.002 34401 15.0V 1.5U 2W 9.001 DISP Connect the 34401A to P4 on the UUT as follows: 9.001 DISP [32] 34401A Input HI to Blue (+) 9.001 DISP [32] 34401A Input LO to Black (-) 9.002 34401 -15.0V 1.5U 2W 10.001 END